我们的网站为什么显示成这样?

可能因为您的浏览器不支持样式,您可以更新您的浏览器到最新版本,以获取对此功能的支持,访问下面的网站,获取关于浏览器的信息:

|本期目录/Table of Contents|

国产室间隔缺损封堵器的动物实验研究(PDF)

《心脏杂志》[ISSN:1009-7236/CN:61-1268/R]

期数:
2005年第2期
页码:
169-171
栏目:
介入治疗
出版日期:
2005-03-05

文章信息/Info

Title:
The experimental study of domestic nitinol ventricular septal defect occluder
作者:
齐静1 朱鲜阳2 王辉山3 姜辉3 宋福林4 侯传举2 全微2 柳阳2
1.大连医科大学,辽宁 大连116027;沈阳军医总医院:2.先心内科,3. 心外科,4.病理科,辽宁 沈阳 110016
Author(s):
QI Jing1 ZHU Xianyang2 WANG Huesan3 JIANGHui3 SONG Fulin4 HOU Chuanju2 QUAN Wei2 LIU Yang2
1. Medical University of Dalian, Dalian, Liaoning 116027, 2.Genetal Hospital of Shenyang Army, Shenyang, Lioaning 110016,China
关键词:
镍钛记忆合金封堵器室间隔缺损疾病模型动物
Keywords:
domestic nitinol occluder ventricular septal defect disease model animal
分类号:
R541.1
DOI:
-
文献标识码:
A
摘要:
目的 观察国产室间隔缺损封堵器通过导管经心室植入动物体内后的封堵效果及组织学反应,评价其可行性、安全性及疗效。方法 选用犬12只,在经心外膜超声的指引,非体外循环下建立室间隔缺损动物模型,直视下经导管植入国产室间隔缺损封堵器。术后15 d、1、3、6个月分别处死动物,进行大体解剖、光镜及电镜观察。结果 8只犬成功建立了室间隔缺损的动物模型并进行了封堵,均存活至观察终点,超声心动图随访无残余分流,生长发育无异常,无严重并发症发生,1~6个月的大体解剖可见封堵器表面被厚度不等的半透明、光滑的新生组织所覆盖,封堵器与室间隔组织结合紧密,光镜及扫描电镜检查证实为内皮细胞、纤维组织。各重要脏器无血栓栓塞及其它异常表现。结论 国产的室间隔缺损封堵器具有良好的组织相容性,用外科方法修补室间隔缺损以及在超声心动图的指引下经心室植入室间隔缺损封堵器是可行的、安全的。
Abstract:
AIM To evaluate the feasibility, safety and efficiency of a domestic nitinol occluder for perventricular closure of ventricular septal defect. METHODS Twelve dogs are prepared for the experiment, after the chest and pericardium are opened, a complete, epicardial echocardiographic assessment of the ventricular septum is performed. A ventricular septal defect animal model was established under echocardiography guidance without exreacorpotreal circulation, the animals were killed for observation at various intervals after successful closure. RESULTS Successful closure was achieved in 8 dogs. Postmortem gross and microscopic examination revealed no dislogement of the device. Gross inspection showed that the devices were covered with a different glistening, smooth, thin layer of endocardium. There was continuity between the endocardium over the device and the surrounding septal wall. Through light microscopy and electron microscopy, fibrous tissue and endothelia cell were seen. The heart, lungs, and great vessels show no evidence of embolization or any other abnormality. CONCLUSION This observation showed the satisfactory biocompatibility, feasibility and safety of the domestic nitinol ventricular septal defect occluder by perventricular closure.

参考文献/References

[1] Lock JE, Block PC, McKay RG, et al. Transcatheter closure of ventricular septal defects[J]. Circulation,1988,78:361-368.

[2] Arora R, Trehan V, Thakur AK,et al. Transcatheter closure of congenital muscular ventricular septal defect[J]. J Interv Cardiol, 2004,17:109-115.

[3] Thanopoulos BD, Tsaousis GS, Karanasios E,et al. Transcatheter closure of perimembranous ventricular septal defects with the Amplatzer asymmetric ventricular septal defect occluder: preliminary experience in children[J]. Heart,2003,89:918-922.

[4] Amin Z, Gu X, Berry JM, et al. New device for closure of muscular ventricular septal defects in a canine model[J]. Circulation, 1999, 100: 320-328.

[5] Gu X, Han YM, Titus JL, et al. Transcatheter closure of membranous ventricular septal defects with a new nitinol prosthesis in a natural swine model[J] . Catheter Cardiovasc Interv, 2000, 50 : 502-509.

[6] Kanlitz R, Panl T, Hauxdorf G, et al. Extending the limits of transcatheter closure of atrial septal defects with the double umbrella device[J]. Heart,1998, 80: 54-59.

[7] Hijazi ZM, Hakim F, Fadley F, et al. Transcather closure of single muscular ventricular septal defects using the Amplatzer muscular VSD occluder: initial results and technical consideration[J]. Catheter Cardiovasc Interv, 2000, 49: 167-172.

[8] Bacha EA, Cao QL, Starr JP, et al. Perventricular device closure of muscular ventricular septal defects on the beating heart: technique and results[J]. J Thorac Cardiovasc Surg, 2003, 126:1718-1723.

备注/Memo

备注/Memo:
收稿日期:2004-12-16.
更新日期/Last Update: 2010-01-05