双源CT 100 kV管电压在冠脉成像中的应用价值

《心脏杂志》[ISSN:1009-7236/CN:61-1268/R]

期数:
2015年第2期
页码:
198-201
栏目:
临床研究
出版日期:
2014-10-25

文章信息/Info

Title:
Application of dual-source CT of 100 kv tube voltage in low-dose computer tomography coronary angiography
作者:
赵国华徐 冰唐萍萍
(北京市顺义区医院放射科,北京 101300)
Author(s):
ZHAO Guo-hua XU Bing TANG Ping-ping
(Department of Radiology, Shunyi District Hospital, Beijing 101300, China)
关键词:
电子计算机体层扫描双源管电压冠状动脉辐射剂量
Keywords:
dual-source CT tube voltage coronary artery radiation dosage
分类号:
R541.4;R445.3
DOI:
-
文献标识码:
A
摘要:
目的:探讨双源CT应用100 kV管电压扫描方案在冠状动脉成像中的应用价值。方法:将70例临床怀疑或确诊为冠心病患者,随机分为管电压100 kV组(低剂量组,35例)与120 kV组(常规组,35例)。其他条件一致,分别对两组图像质量和辐射剂量进行评价。入组条件:体质量指数(BMI)为20~24,窦性心率。结果:两组冠脉成像图像质量无显著差别,而100 kV辐射剂量较120 kV组显著降低(P<0.01)。结论:应用双源CT做冠脉CT血管造影,给予100 kV管电压能够在保证图像质量,满足诊断要求同时降低辐射剂量。
Abstract:
AIM:To evaluate the application of dual-source CT 100 kv tube voltage in low-dose computer tomography coronary angiography. METHODS: Seventy patients suspected of or diagnosed as having coronary disease (with BMI 20-24% and sinus rhythm) were randomly divided into two equal groups: study group (100 kv tube voltage) and control group (120 kv tube voltage). The image quality and radiation dose of the two groups were compared. RESULTS: No statistical difference was found in the scores of imaging quality between the two groups. The average effective radiation dose of the study group and control group were (3.29±0.56) mSv and (7.22±1.43) mSv, respectively, with significant differences between groups. CONCLUSION: Application of 100 kv tube voltage in CT coronary angiography can ensure good image quality and high diagnostic accuracy of coronary artery stenosis while it significantly reduces the radiation dose.

参考文献/References

[1]Austen WG,Edwards JE,Frye RL,et al.A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association[J].Circulation,1975,51(4 suppl):5-40
[2]Earls JP,Berman EL,Urban BA,et al.Prospectively gated transverse coronary CT angiography versus retrospective gated helical technique:improved image quality and reduce radiation dose[J].Radiology,2008,246(3):742-753.
[3]Comission of the Equropean Communities.European guidelines on quality criteria for computed tomography;EUR 16262EN[M].Luxembourg:Office for the official publications of th European commission,2000.
[4]Mollet NR,Cademartiri F,Van Mieghem CA,et al.Hight-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography[J].Circulation,2005,112(10):2318-2323.
[5]Szucs-Farkas Z,Kurman L,Strautz T,et al.Patient exposure and image quality of low-dose pulmonary compute tomography angiography:comparison of 100- and 80-kvp protocols[J].Invest Radiol,2008,43(12):871-876.
[6]Leschka S,Stolzmann P,Desbiolles L,et al.Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses:first experience[J].European radiology,2009,19(12):2896-2903.
[7]Golding SJ,Shrimton PC.Radiation dose in CT:are we meeting challenge?[J].Br Radiol,2002,75(1):1-4.
[8]Dixon AK,Goldstone KE.Abdominal CT and the euratom directive[J].Eur Radiol,2002,12(6):1567-1570.

备注/Memo

备注/Memo:
收稿日期:2014-05-31.
作者简介:赵国华,主治医师 Email:changj3401@163.com
更新日期/Last Update: 2014-11-18