海尔BP3BG1-A全自动上臂电子血压计准确性的临床评价

《心脏杂志》[ISSN:1009-7236/CN:61-1268/R]

期数:
2015年第1期
页码:
92-094
栏目:
临床研究
出版日期:
2014-09-25

文章信息/Info

Title:
Validation of Haier BP3BG1-A automatic upper arm blood pressure monitor for clinical use and self-measurement
作者:
王 勇1薛玉生12王 彬2
(1.柞水县中医医院内科, 陕西 商洛 711400;
2.第四军医大学附属唐都医院心内科,陕西 西安 710038)
Author(s):
WANG Yong1 XUE Yu-sheng2 WANG Bin2
(1.Department of Medicine, Hospital of Traditional Chinese Medicine, Zhashui County, Shangluo 711400, Shaanxi, China;
2.Department of Cardiology, Tangdu Hospital, Fourth Military Medical University, Xi’an 710038, Shaanxi, China)
关键词:
血压欧洲高血压协会国际协议验证
Keywords:
blood pressure European Society of Hypertension International Protocol validation
分类号:
R540.4
DOI:
-
文献标识码:
A
摘要:
目的:按照《欧洲高血压协会国际质量控制方案2010修正版(ESH-IP 2010)》,评估海尔BP3BG1-A全自动上臂电子血压计临床应用的准确性。方法: 测量者采用水银血压计和受试血压计先后测量33名被试者同一左侧上臂收缩压和舒张压。根据ESH-IP 2010,每位被试者测量3对血压,共收集99例次受试血压计与水银血压计的测值。结果: 受试血压计测得73次(74%)收缩压与水银血压计测值差在5 mmHg以内,98次(99%)在10 mmHg以内,99次(100%)在15 mmHg以内;86次(87%)舒张压与水银血压计测值差5 mmHg以内,98次(99%)差10 mmHg。受试血压计测得收缩压与水银血压计测值差为(3.1±3.7) mmHg;舒张压测值差为(-0.9±3.7) mmHg。采用受试血压计与水银血压计的测值2~3次的差别都在5 mmHg以内的受试者数目分别为26名(79%)和29名(88%)。结论: 受试血压计对临床成人患者血压测量的准确性达到《 ESH-IP 2010》的验证标准。
Abstract:
AIM:To validate the Haier BP3BG1-A automatic upper arm blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. METHODS: Sequential same-left-arm measurements of systolic and diastolic blood pressure were obtained in 33 participants using the mercury sphygmomanometer and the test device. According to the validation protocol, 99 pairs of the test device and reference blood pressure measurements (three pairs for each of the 33 participants) were obtained in the study. RESULTS: The device produced 73, 98 and 99 measurements within 5, 10 and 15 mmHg for systolic blood pressure (SBP) and 86, 98 and 99 for diastolic blood pressure (DBP), respectively. The mean SD device-observer difference was (3.1±3.7) mmHg for SBP and (-0.9±3.7) mmHg for DBP. The number of subjects with two or three of the device-observer difference within 5 mmHg was 26 for SBP and 29 for DBP. CONCLUSION: Haier BP3BG1-A automatic upper arm blood pressure monitor can be recommended for clinical use and self-measurement in an adult population on the basis of the ESH-IP revision 2010.

参考文献/References

[1]He J,Whelton PK.Elevated systolic blood pressure and risk of cardiovascular and renal disease:Overview of evidence from observational epidemiologic studies and randomized controlled trials[J].Am Heart J,1999,138(3 Pt 2):211-219.
[2]Klag MJ,Whelton PK,Randall BL,et al.Blood pressure and end-stage renal disease in men[J].N Engl J Med,1996, 334(1):13-18.
[3]Blood Pressure,Cholesterol,and stroke in eastern Asia Eastern Stroke and Coronary Heart Disease Collaborative Research Group[J].Lancet,1998,352(9143):1801-1807.
[4]Kearney PM,Whelton M,Reynolds K,et al.Worldwide prevalence of hypertension:a systematic review[J].J Hypertens,2004,22(1):11-19.
[5]Cappuccio FP,Kerry SM,Forbes L,et al.Blood pressure control by home monitoring:meta-analysis of randomised trials[J].BMJ,2004,329(7458):145.
[6]O’brien E,Pickering T,Asmar R,et al.Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults[J].Blood Press Monit,2002,7(1):3-17.
[7]O’brien E,Petrie J,Littler WA,et al.The British hypertension society protocol for the evaluation of blood pressure measuring devices[J].J Hypertens,1993,11(Suppl 2):S43-S63.br />[8]O’Brien E,Atkins N,Stergiou G,et al.European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults[J].Blood Press Monit,2010,15(1):23-38.
[9]Friedman BA,Alpert BS,Osborn D,et al.Assessment of the validation of blood pressure monitors:a statistical reappraisal[J].Blood Press Monit,2008,13(4):187-191.
[10]Wan Y,Heneghan C,Stevens R,et al.Determining which automatic digital blood pressure device performs adequately:a systematic review[J].J Hum Hypertens,2010,24(7):431-438.

备注/Memo

备注/Memo:
收稿日期:2014-05-01.
通讯作者:薛玉生,主任医师,主要从事介入性心脏病学研究Email:xys978088@163.com
作者简介:王勇,主治医师Email:2442902591@qq.com
更新日期/Last Update: 2015-01-20